Ploidy Installation Rev. 02

Major VersionM

by Pete

  1. The Ploidy Analyzer is shipped in a cardboard box that is attached to a wooden pallet with a plastic wrap. Remove the wrap and open the top of the box with a knife or razor blade. Remove foam packing, inserts, bottles and cap assemblies (sheath and waste).
    • The Ploidy Analyzer is shipped in a cardboard box that is attached to a wooden pallet with a plastic wrap.

    • Remove the wrap and open the top of the box with a knife or razor blade.

    • Remove foam packing, inserts, bottles and cap assemblies (sheath and waste).

  2. Observe fluidic connections at rear of instrument (labeled waste, sheath and air) from left to right. Factory ships with knotted tubing on connections. Remove knotted tubing from fittings. Keep knotted tubing connectors for future transport.
    • Observe fluidic connections at rear of instrument (labeled waste, sheath and air) from left to right. Factory ships with knotted tubing on connections.

    • Remove knotted tubing from fittings.

    • Keep knotted tubing connectors for future transport.

    • Unscrew blue plastic collars from threaded fittings, place collar on end of tubing of sheath and waste connections, attach respective tubing to fitting and tighten collars hand tight.'''

  3. Insert waste level sensor plug into connector labeled "Waste".
    • Insert waste level sensor plug into connector labeled "Waste".

    • Insert sheath level sensor plug into connector labeled "Sheath".

    • Careful not to swap "Sheath with Waste"

  4. To reduce bubbles angle bottle while filling and have sheath run down inside of bottle. Fill sheath bottle to approximately 700ml mark. When inserting probe and pickup tube assembly move the assembly up and down several times to reduce likelihood of trapping a bubble in the sheath line prior to attaching and tightening the cap.
    • To reduce bubbles angle bottle while filling and have sheath run down inside of bottle.

    • Fill sheath bottle to approximately 700ml mark.

    • When inserting probe and pickup tube assembly move the assembly up and down several times to reduce likelihood of trapping a bubble in the sheath line prior to attaching and tightening the cap.

    • Attach waste cap assembly to waste bottle. Secure tightly by hand but do not overtighten and crack the cap. This bottle acts as a vacuum reservoir since it is a vacuum based system.

  5. Insert end of electrical cord below master power switch and check all fluid and sensor connections. Plug the other end of the power cord into suitable wall outlet or PowerVar UPS if available. Plug the keyboard and mouse USB connections into the computer connections located at the bottom left side of the instrument.
    • Insert end of electrical cord below master power switch and check all fluid and sensor connections. Plug the other end of the power cord into suitable wall outlet or PowerVar UPS if available.

    • Plug the keyboard and mouse USB connections into the computer connections located at the bottom left side of the instrument.

    • Locate main cover access screws and remove with a 5mm hex wrench

    • Video showing how How to link to video.

  6. Provide support under the left side of the cover when opened so as not to put excessive stress on the hinge support. For support, a box of lens cleaners or chem wipes, or test tube rack can be used.
    • Provide support under the left side of the cover when opened so as not to put excessive stress on the hinge support.

    • For support, a box of lens cleaners or chem wipes, or test tube rack can be used.

    • Main cover can now be lifted and moved to the left on its hinge point.

  7. Upon opening the cover you will notice foam packing on the optical bench, this needs to be removed as part of the setup. Please save and reuse if the instrument will be transported
    • Upon opening the cover you will notice foam packing on the optical bench, this needs to be removed as part of the setup. Please save and reuse if the instrument will be transported

    • If the instrument will be aligned see the Alignment Procedure for insertion of interlock key, if not the instrument cover can be closed and secured with the screws.

    • Remove the sample tube from the sample port if present.

    • Insert and tighten both long 5mm hex head screws.

    • If you are having difficulty with screw insertion see videos on steps 10 and 11.

    • See video for location of conical guides for screw threads.

    • See video for tips and tricks for the securing screws.

  8. Ensure the master switch (located at the rear above the power cord receptacle) is flipped to the ON position. Remove the protective paper if it is present. Turn the instrument on by pushing the power button located at the left rear of top cover (main power switch at rear of instrument needs to be on as well)
    • Ensure the master switch (located at the rear above the power cord receptacle) is flipped to the ON position.

    • Remove the protective paper if it is present.

    • Turn the instrument on by pushing the power button located at the left rear of top cover (main power switch at rear of instrument needs to be on as well)

    • The system will take a few minutes to completely boot up to the Windows desktop.

  9. Double click on the Cube 16 icon (Cube 15 if it is an older system) to open up the CyView application. Note, some systems may automatically open into the CyView application upon bootup. At the login screen left click in the Login area and type USER, click in the Password area and type Cube1. You can either press the Enter key or left click on Work with CyView.
    • Double click on the Cube 16 icon (Cube 15 if it is an older system) to open up the CyView application. Note, some systems may automatically open into the CyView application upon bootup.

    • At the login screen left click in the Login area and type USER, click in the Password area and type Cube1. You can either press the Enter key or left click on Work with CyView.

  10. Select the Prime function by left clicking on the Prime icon and start the Prime sequence by left clicking on the arrow located at the bottom right (labeled as Start the Selected Batch if the mouse is over it. Select the Prime function by left clicking on the Prime icon and start the Prime sequence by left clicking on the arrow located at the bottom right (labeled as Start the Selected Batch if the mouse is over it.
    • Select the Prime function by left clicking on the Prime icon and start the Prime sequence by left clicking on the arrow located at the bottom right (labeled as Start the Selected Batch if the mouse is over it.

  11. Follow the onscreen instructions and place a 12 x 55 mm tube half filled with the Decontamination Solution (item 04-4010_R) up into the sample port until it clicks. On screen prompts will indicate what needs to be done throughout the Prime process. Proceed accordingly and use a tube half filled with DI H2O at the last step as indicated.
    • Follow the onscreen instructions and place a 12 x 55 mm tube half filled with the Decontamination Solution (item 04-4010_R) up into the sample port until it clicks.

    • On screen prompts will indicate what needs to be done throughout the Prime process.

    • Proceed accordingly and use a tube half filled with DI H2O at the last step as indicated.

  12. Left click on the Work icon (to the right of the Prime icon) and
    • Left click on the Work icon (to the right of the Prime icon) and

    • Left click the Start the Batch arrow to put the instrument in the standard run mode.

    • At this point the instrument is ready for running the calibration (alignment beads). Depending on the configuration you will need to check the alignment for the UV LED, Green Laser or both.